|












| |
ThermoProbe
parts & accessories
Service & parts
ICL is a factory authorized representative,
service, parts and calibration facility. We stock all parts for all current
ThermoProbe models and can repair and re-calibrate (with new certificate) all models
quickly, professionally and at competitive rates.
We have "loaner" ThermoProbes
available for emergency situations. Call us by 4:00 pm Eastern time and you'll have
a loaner the next morning! There is no charge for the "loaner" service,
except for shipping. Just send us back the "loaner" when you get your
repaired ThermoProbe back from us.
Some commonly needed
"user-replaceable" probe & cable assemblies: (Please be sure to refer
to our
instructions for changing probe & cable
assemblies and calibration of your ThermoProbe!)
| Part Number |
Description |
| PG-050-SW-SM |
Replacement sensor & cable
assembly for TP-2C, TP-5C, TP-7, TP-7C, TP-8 or TP-9
50 foot (15m) length, marked at 5 foot increments with brass ferrules
Standard duty |
| PG-075-SW-SM |
Replacement sensor &
cable assembly for TP-2C, TP-5C, TP-7, TP-7C, TP-8 or TP-9
75 foot (23m) length, marked at 5 foot increments with brass ferrules
Standard duty |
| PG-110-SW-SM |
Replacement sensor &
cable assembly for TP-2C, TP-5C, TP-7, TP-7C, TP-8 or TP-9
110 foot (33m) length, marked at 5 foot increments with brass ferrules
Standard duty |
| PG-075-AW-SM |
Asphalt weight
replacement sensor & cable assembly for TP-2C, TP-5C, TP-7, TP-7C, TP-8 or
TP-9
75 foot (23m) length, marked at 5 foot increments with brass ferrules
Extra heavy weighted sensor and heavy duty cable |
These are only a few of the many cables
we have IN STOCK for all the ThermoProbe models!

When you change the sensor/cable assembly you
must adjust your ThermoProbe to the particular characteristics of the new RTD sensor.
You cannot just physically change out the cable and go! You will have
severe errors in reading! Please see our
instructions
for changing probe & cable assemblies and calibration of your ThermoProbe!
HOW OFTEN DO YOU NEED TO CALIBRATE (OR
CHECK) YOUR THERMOPROBE?
The American Petroleum Institute's (API)
publication 'Manual of Petroleum Measurement Standards' (Chapter 7, Section 3) 'Static
Temperature Determination Using Portable Electronic Thermometers' (First Edition, July
1985) offers specific recommendations for recalibration and verification of your
electronic digital thermometer. The following section, in brown type, has been reprinted,
with permission, courtesy of the American Petroleum Institute, as a customer service.
The newest version of Chapter 7, dated June 2001
has the same requirements.
To purchase this or other API
standards, please contact the American Petroleum Institute's Publications Department at
202-682-8000
7.8.2 Initial Calibration
Before initial use, and at least once a year thereafter, each portable electronic
thermometer shall be re-standardized in a laboratory or other qualified calibration
facility. The thermometer shall be re-standardized by comparing it at three or more
temperatures with either a National Bureau of Standards (NBS)
(ICL
note: NBS is now NIST)
certified thermometer or an
equivalent thermometer with accuracy traceable to the NBS in accordance with the
manufacturer's instructions. These standardization checks will ensure that accuracy is
maintained within the limits given in Table 3.
8.2.1 Field verification
Before each use, or once per day (whichever is less frequent), portable electronic
thermometers should be spot checked by comparing the ambient reading against an ASTM glass
stem thermometer in liquid. If the readings differ by more than 0.5 degree F (or 0.25 degree
C), the portable electronic thermometer should be re-standardized before it is used for
custody transfer.
8.2.2 MONTHLY VERIFICATION AND
INSPECTION
On a monthly schedule, the portable electronic thermometer should be checked at two or
more temperatures near the ends of its range against an NIST certified thermometer or an
equivalent thermometer with accuracy traceable to the NIST. Make this check by placing the
two thermometers side-by-side in a circulating bath and leaving them undisturbed for at
least 10 minutes before making comparative readings. Alternately, the ice point and
boiling point of fresh water may be used for accuracy check. The boiling point of fresh
water must be corrected for altitude. If the thermometer readings differ by more than the
tolerances listed in Table 1, the thermometer should be re-standardized (see 7.8.2)
In addition, the following physical
checks should be made:
1. The junction between the cable and the probe should be checked for mechanical damage.
2. The cable insulation should be checked for cuts, breaks, or abrasion.
|
TABLE 3 - Portable Electronic Gauging Thermometers |
|
Graduation |
Range |
Required accuracy |
|
0.1 F |
0-200oF |
+/- 0.2oF |
|
|
> 200oF |
+/- 0.5oF |
|
0.1 C |
0-100oC |
+/- 0.1oC |
|
|
> 100oC |
+/-
0.3oC |
| Reproduced
with permission from API |
ICL Calibration Laboratories can perform this
calibration and certification for you with a one to two day turnaround.
More
ThermoProbe
models ThermoProbe sensor/cable
replacement & calibration instructions
|